The account reported finding missing the round edge, and the complaint stated that the tip seems to be missing the round edge resulting in jagged edge which is dangerous for the patient.A photo was received for our investigation, visual inspection was conducted with the provided pictures, and finished good production records were reviewed.Based on the complaint details, picture evaluation and our investigation, the complaint was confirmed.The vendor identified that during the production process, an abnormal alarm was triggered by the equipment and a machine adjustment action was initiated; after the machine adjustment, the products were not specifically distinguished and monitored, resulting in not being selected in full product inspection, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|