It was reported that during a da vinci-assisted surgical procedure, customer noticed sparks coming from the permanent cautery spatula instrument insulation of the electrocautery tip while clearing pelvic lymph nodes.The procedure was completed with no reported injury.The arcing appear to originate/occur from equipment, wrists, insulating materials and places.The arc grounded on patient anatomy when the monopolar coag energy was activated.Burnt insulation was noticed on the instrument.The instrument tips did not collide with any other instrument or tool during procedure.It was reported that there was no injury to the patient.The procedure was completed as planned.
|
|
The permanent cautery spatula instrument has not been received for investigation.The video of this procedure was reviewed.From the video alone, it is not possible to determine if any arcing, sparks or burnt insulation were present.In the video, the operator is dissecting tissue using the monopolar cautery spatula while retracting locally with the fenestrated forceps.The tip of the spatula is removed from tissue contact then the wrist of the instrument rests against the tissue where some heat appears to be generated along with smoke.The wrist is rolled, making tissue contact again, generating heat and smoke once again.No arcing was confirmed from the video.
|