The reported issue was tubing was kinking.A review of the device history records was performed for the reported packaging material and component, and the review confirms the batches met all material, assembly and packaging specifications; the samples were not returned to medline, but the customer did provide photos of the reported issues.Based on the photos provided by the customer, the product exhibits some flat spots and dents in the fluid deliver sets, and the complaint is not confirmed for kinks.The tubing can develop dents or flattened spots as it rests against itself or other components within the pouch; dhr review revealed no quality related issues or manufacturing deficiencies at the time of manufacture, and based on no adverse trends, no issues noted in the dhr, and adequate process controls in place, no additional corrective actions to be taken by medline at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|