The facility reported that on (b)(6) 2024, the nebulizer gets warm and doesn¿t dispense the medication; the user did not miss a treatment, and no medical intervention, serious injury, or follow-up care has been reported.Medline quality received one unit, visually inspected the sample, observed that the sample was received with a used filter, plugged the sample into a power source, and the sample turned on and misted; the digital flowmeter reading was 7.9 lpm, the unit arrived in good condition with all wires and tubing properly connected, the unit ran three times at 30 minute intervals without issue, and the average temperature was 88 degrees fahrenheit through the vent and 100 degrees fahrenheit with the cover off, which meets product specifications.The reported problem was not confirmed with the received sample, and medline was not able to confirm what may have caused the issue.The quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|