Catalog Number INT413
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 05/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reports that the dental implants failed due to peri-implantitis at tooth sites 16 and 17.The doctor reports in the per that the procedure was not concluded by placing another implants.Site grafted.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4) a1: patient identifier: unknown / not provided a2: age at time of the event: unknown / not provided a4: patient weight: unknown / not provided d10: concomitant medical products: int413, osseotite® tapered certain® implant 4 x 13mm, 2021020282.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® IMPLANT 4 X 13MM
Common Device Name
Dental Implant
Manufacturer (Section D)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZimVie US Corp LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
+1(561)971-3230
MDR Report Key50931627
Report Number0001038806-2026-04950
Device Sequence Number9498625
Product Code DZE
UDI-Device Identifier00844868006664
UDI-Public(01)00844868006664(17)260217(10)2021020282
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K063341
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/17/2026
Device Catalogue NumberINT413
Device Lot Number2021020282
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/10/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/10/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Dental Implants See H11 Narrative
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-