The account reported that during routine intubation for surgery, placement of a lubricated endotracheal tube resulted in minor glottic bleeding due to a hard sharp end of the endotracheal tube, and this was more than anticipated given the ease of intubation.No sample or photo was received for investigation, finished good production records could not be reviewed since a lot number was not provided, and component trending was reviewed with 0 additional reported complaints for this issue for the last 24 months.Based on the complaint details and investigation, the complaint is not confirmed, and the investigation was limited without receipt of the lot number, a physical sample, or detailed photographic evidence from the customer.The sample was required to determine cause, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|