| Catalog Number |
02K41-27 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed a falsely elevated architect stat troponin-i result for a 15-year-old female patient.The following data was provided (normal range <0.04): the initial architect stat troponin-i result for sid (b)(6) was 0.786, rerun 0.000.The sample was repeated twice on another instrument, and the results were 0.004 and 0.002.No impact on patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Data and information provided by the customer were reviewed and support the complaint issue.A ticket search by lot did not identify any trends for lot 42171un25.A review of tracking and trending for the architect stat troponin-i did not identify any trends.A labeling review concludes that the complaint issue is adequately addressed.A device history record review did not identify any nonconformances, potential nonconformances or deviations associated with the complaint lot.Customer field data was used to assess the performance of the architect stat troponin-i using worldwide data.The patient median data and calibrator f rlu mean were analyzed and compared to an established control limit.This evaluation indicated that the patient medians and the calibrator f rlu mean for the lot for lot 42171un25 are within the established limits.Therefore, no unusual reagent lot performance was identified for lot 42171un25.Based on our investigation, no systemic issue or deficiency with the architect stat troponin-i lot 42171un25 was identified.
|
| |
|
Search Alerts/Recalls
|