| Model Number |
PCF-H190DL |
| Medical Device Problem Codes |
Residue After Decontamination (2325); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device the initial evaluation findings were confirmed.The root cause of the reportable malfunction, no illumination from the light guide lens, was traced to component failure.However, the most likely cause of the issue involving white residue within the air/water channel and air/water cylinder could not be established.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer returned the colonovideoscope for a separate issue.During the initial evaluation of the returned device, white residue was observed within the air/water channel and air/water cylinder, and no illumination was observed from the light guide lens.There was no patient involvement.
|
| |
|