The account reported finding linting towels.A photo and a sample were received for investigation; 4 unused folded towels were received, two or towels had linting and two did not, and finished good production records and finished good pack build were reviewed.The reported issue was confirmed, as the two or towels found linting had a chunk of towel lint on the inside of the or towels.Based on the complaint details the root cause is considered to be a vendor product issue relating to the component manufacturing process, and the vendor response stated the root cause may be related to the sewing process if the suction outlet on the equipment is blocked; preventive measures include cleaning the tuyere of the sewing machine and training operators to clean the equipment according to requirements.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|