| Model Number |
e 411 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys estradiol iii assay results for 2 patient samples on a cobas e 411 analyzer (disk system).Patient 1's initial result was 680.8 pg/ml.The repeat results were 51.72 pg/ml and 61.05 pg/ml.Patient 2's initial result on 09-sep-2026 was 166.7 pg/ml.The repeat result was 52.57 pg/ml.The repeat results were deemed to be correct, as the initial high results did not match clinical expectations.
|
| |
|
Additional Manufacturer Narrative
|
|
The elecsys estradiol iii assay reagent lot number is 97554301, and the expiration date is 30-jun-2027.Qc was within range at the time of the event on 07-sep-2026.The system was released for sample measurement.The qc level 2 was out of range on 08-sep-2026.Qc results for 09-sep-2026 were not available.A field service engineer (fse) determined that the z-motor of the pipetting/reagent needle was the root cause.The fse disassembled all of the covers, cleaned all surfaces to remove dust, and replaced the z-motor of the pipetting/reagent needle.For verification, calibration runs, and patient sample processing were performed without any issues.The analyzer is working according to specification after the service.The investigation determined that the service actions resolved the issue.
|
| |
|