The customer reported no alarm after reading is done and also stated it is not beeping and he held the button for 180 seconds and did not hear a beep.Medline received two digital thermometers in used condition with working battery, sanitized the samples, tested them according to the instructions for use, and took two different measurements per thermometer.The complaint was not confirmed; the signaling tone was noticeable after the lo sign was displayed, heard at the end of every test, and audible when the samples were turned off.The cause was documented as not confirmed with received sample, and the sample worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|