| Model Number |
V60 V60Plus Ventilator |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported that the v60 ventilator stand had a broken locking caster.There was no patient involvement when the issue occurred.No patient impact and/or harm was reported.No user impact and/or harm was reported.A philips remote service engineer (rse) provided assistance to the bme, and it was reported that the issue was replicated during diagnostics.The bme was provided with the part numbers (locking casters, and caster wrench) for replacement parts.The issue was replicated/confirmed during diagnostics.Based on the details provided, the damaged castors are considered a malfunction and can become a stability issue for the device.This investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
A follow up was performed with the customer, and it was reported that the casters were replaced to resolve the issue.The device was reported to have been returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.The cause of the broken locking caster could not be determined; the customer did not provide any further details regarding the event.A conclusion/root cause cannot be provided, as the suspected part has not been returned for further analysis.
|
| |
|