| Model Number |
78768-01 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The customer received sensor scan results of 69.00 mg/dl compared to readings of 202.00 mg/dl respectively obtained on a meter and the results, when plotted on a parkes error grid, fall into the c zone, showing the difference in values to be clinically significant.The length of sensor wear was 14 days.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The product has been requested back for investigation.A follow-up report will be submitted upon completion of investigation activities or if additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|