The customer stated that, when the patient went to raise the head of the bed, he heard a crack and the bed would not work by remote; no injury occurred.A photo was provided, and after reviewing the photo it was discovered that part of the motor box casing appears to be completely fractured off, the head section of the motor box appears to be fractured, and the head actuator appears to be exposed due to the fracture; functionality of the motor box can not be determined with photos.Investigation results: confirmed.Given the evidence provided in the photo the root cause of this issue is unknown; the division will monitor through trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|