MAUDE Adverse Event Report: Insulet Corporation Omnipod 5 Controller; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Insulet Corporation Omnipod 5 Controller; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number PT-001827
Medical Device Problem Codes Overheating of Device (1437); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the patient that the omnipod 5 controller/personal diabetes manager (pdm) will no longer hold a charge, became excessively hot and while charging; there was melting and bulging near the charging port.The patient also reported that the controller was struggling to stay connected to their dexcom g7 and was slowly losing charge.The patient stated that they were using a black charging cord.No harm to the patient was reported and no medical assistance was required.A replacement pdm was issued to the patient.
 
Additional Manufacturer Narrative
The device has been received; however, an investigation has not yet been completed.When the investigation is complete, a supplemental report will be submitted with the investigation results of the reported thermal event.The reported thermal device malfunction is associated with a personal diabetes manager manufactured after the correction for action res#91127 was implemented.Lot release records were reviewed, and the product lot met all acceptance criteria.Locked down smartphone: lockdown omnipod software app version: 4.0.2 operating system: n5004l-am-q-mv01602-06-01.06 hardware: n5004l cgm sensor type: g7 please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Omnipod 5 Controller
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer (Section G)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
+1(978)600-7000
MDR Report Key50932074
Report Number3014585508-2026-71945
Device Sequence Number9589120
Product Code QFG
UDI-Device Identifier10385083000022
UDI-Public(01)10385083000022(11)230329(10)H000509
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K203768
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPT-001827
Device Catalogue NumberPDM-H001-G-XX
Device Lot NumberH000509
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/14/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured03/29/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-