| Model Number |
PT-001827 |
| Medical Device Problem Codes |
Overheating of Device (1437); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the patient that the omnipod 5 controller/personal diabetes manager (pdm) will no longer hold a charge, became excessively hot and while charging; there was melting and bulging near the charging port.The patient also reported that the controller was struggling to stay connected to their dexcom g7 and was slowly losing charge.The patient stated that they were using a black charging cord.No harm to the patient was reported and no medical assistance was required.A replacement pdm was issued to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received; however, an investigation has not yet been completed.When the investigation is complete, a supplemental report will be submitted with the investigation results of the reported thermal event.The reported thermal device malfunction is associated with a personal diabetes manager manufactured after the correction for action res#91127 was implemented.Lot release records were reviewed, and the product lot met all acceptance criteria.Locked down smartphone: lockdown
omnipod software app version: 4.0.2
operating system: n5004l-am-q-mv01602-06-01.06
hardware: n5004l
cgm sensor type: g7
please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|
Search Alerts/Recalls
|