• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; Permanent Defibrillator Electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; Permanent Defibrillator Electrodes Back to Search Results
Model Number 7121Q/65
Medical Device Problem Codes Break (1069); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Low impedance (2285); Charging Problem (2892)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2026
Type of Reportable Event Serious Injury
Event or Problem Description
During a clinic follow-up, a decrease in the defibrillation and pacing impedance and episodes of noise resulting in oversensing and inappropriate anti-tachycardia pacing (atp) were observed on the rv lead.Additionally, an over current detection (ocd) alert was observed.Rv lead damage was alleged, but was unable to be confirmed.Technical support was contacted and the rv lead was capped and replaced to resolve the event.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Common Device Name
Permanent Defibrillator Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50932224
Report Number2017865-2026-20423
Device Sequence Number10725391
Product Code NVY
UDI-Device Identifier05414734503228
UDI-Public(01)05414734503228(10)3508703(17)140930
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2014
Device Model Number7121Q/65
Device Lot Number3508703
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/28/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-