• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; METER,GLUCOSE,EVENCARE PROVIEW,PROF.USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; METER,GLUCOSE,EVENCARE PROVIEW,PROF.USE Back to Search Results
Catalog Number MPH4540
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "meter not reading accurately.".
 
Additional Manufacturer Narrative
The reported issue was "readings inaccurate" and "meter not reading accurately." medline received 1ea blood glucose monitor, performed visual inspection, noted no cosmetic defects nor any damage, replaced the batteries with control batteries on site, and initiated testing utilizing glucose strips and test solution; the device had readings of 73 (68-92) and 242 (216-292).As a result, the complaint could not be confirmed.No further action was taken because objectionable trends were not seen for this issue, and the division was informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER,GLUCOSE,EVENCARE PROVIEW,PROF.USE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
shravan ingole
three lakes drive
northfield, IL 60093--275
MDR Report Key50932285
Report Number1417592-2026-53037
Device Sequence Number10722294
Product Code CGA
UDI-Device Identifier10888277409163
UDI-Public10888277409163
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K111456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH4540
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/01/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2023
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-