A review of the complaint history for (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for (b)(6) was performed from the date of manufacture, 09-apr-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device's universal serial bus (usb) required disabled.A technical support specialist (tss) reviewed the transaction records in medbank myqlink (mql) and confirmed that the medication issue transaction had been successfully processed.The tss then remotely accessed the cabinet, disabled the universal serial bus (usb) power management settings, and rebooted the cabinet.Using the tester application, the tss verified that the lid opened correctly.Further investigation determined that medication packaging material had caused the lid to become stuck.After the obstruction was removed and the cabinet was rebooted, the user successfully completed a cycle count and accessed the medication as expected.The system functioned as intended after the technical support specialist troubleshot the device.
|