The account reported finding adhesive issue dressing, including dressing backing failing and some kits not having the dressing.A photo was received for the investigation, finished good production records were reviewed, and the finished good pack build was reviewed; the photo showed the dressing and the paper frame was detached from the dressing itself so the dressing was stuck to the liner backing.Per the investigation and photo received, the reported issue was confirmed.The cause was identified as deriving from the raw materials process when the force between frame liner and the film is too weak; medline is working with the component manufacturing facility for further evaluation and corrective actions.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|