| Model Number |
7842 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
Asystole (4442)
|
| Date of Event |
01/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited noisy signals due to unexpected programming.Additionally, a red alert was found for high out of range pace impedance measurements.Technical services (ts) was consulted and found the system to be operating in unipolar rather than bipolar with no previous history of a lead safety switch.Ts believed a bigger lead anomaly was present but have no confirmation at this time.Ts noted that pacing inhibition greater than 2 seconds had occurred at times.No additional adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|