| Model Number |
PT-002516 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient's blood glucose level rose to over 400 mg/dl while wearing the pod between 4 and 24 hours.The patient reported the pod was leaking and being unsure if the cannula was properly seated in the infusion site (arm).As treatment, a new pod was applied and a bolus was administered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned; however, the investigation is not yet complete.When the investigation is complete, a supplemental report will be submitted with the investigation results.Lot release records were reviewed and the product lot met all acceptance criteria.Locked down smartphone: lockdown.Omnipod software app version: 4.0.2.Operating system: n5004l-am-q-mv01602-06-01.06.Hardware: n5004l.Cgm sensor type: abbott freestyle libre 2 plus.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|