| Model Number |
500004502000 |
| Medical Device Problem Codes |
Break (1069); Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medflex 1000 the keyboard malfunctioned.As per part investigation, it was determined that the complaint was attributed to the physical damage of part pn 358591-01, which resulted in thermal damage and produced a functional failure that prevented proper keyboard functionality.However, there were no delays or adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 07-feb-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Part analysis:
the report condition of keyboard malfunction was confirmed during fse testing and subsequently confirmed in the dchu inspection process.According to work order # (b)(4), the field service engineer (fse) reported a keyboard cable frayed so the fse replaced the keyboard and the customer verified functionality.During dchu visual inspection:
pn 358591-01: signs of a frayed cable with signs of thermal damage were observed on the visual inspection.During dchu testing:
pn 358591-01: no dchu testing was required due to the frayed cable with signs of thermal damage observed during the visual inspection.The device was in use for treatment purposes as intended per 21 cfr 820.198(d).Root cause:
the root cause of the reported failure keyboard malfunction is attributed to the physical damage of part pn 358591-01 which ultimately caused a thermal damage and produced a functional failure that prevents the proper keyboards functionality.
|
| |
|