• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Philips Medizin Systeme Böblingen GmbH Patient Information Center iX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Philips Medizin Systeme Böblingen GmbH Patient Information Center iX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) Back to Search Results
Model Number Patient Information Center iX
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the central station does not have audible alarms.The device was in use on a patient at the time of the event.There was no adverse event reported.
 
Additional Manufacturer Narrative
Analysis was performed remotely by service personnel and included review of the reported condition, confirmation that no error codes or messages were displayed, efforts to reproduce the issue, and physical inspection of the speaker connection and the connection at the back of the pc.During that analysis the customers monitor technician was asked to check the physical connection of the external speaker on the back of the pc.It was observed that the connection was loose.After the monitor technician reseated the speaker connection and tightening the cable, all alarms were able to be heard.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was a loose speaker connection.The issue was resolved by reseating the external speaker connection and tightening the cable.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Patient Information Center iX
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
+49(703)14632031
MDR Report Key50933053
Report Number9610816-2026-102051
Device Sequence Number21522473
Product Code MHX
UDI-Device Identifier00884838132320
UDI-Public(01)00884838132320(10)4.8.2.0
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K211900
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPatient Information Center iX
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/17/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-