Analysis was performed remotely by service personnel and included review of the reported condition, confirmation that no error codes or messages were displayed, efforts to reproduce the issue, and physical inspection of the speaker connection and the connection at the back of the pc.During that analysis the customers monitor technician was asked to check the physical connection of the external speaker on the back of the pc.It was observed that the connection was loose.After the monitor technician reseated the speaker connection and tightening the cable, all alarms were able to be heard.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was a loose speaker connection.The issue was resolved by reseating the external speaker connection and tightening the cable.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|