| Model Number |
GIF-HQ190 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the gastrointestinal videoscope to the service center for a separate issue.During the initial evaluation of the device, there was an e217 error (scope error).There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.The device evaluation found there was an e217 error (scope error).Based on the results of the investigation, the probable root cause was component failure, the printed circuit board was defective.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post market surveillance activities.
|
| |
|