Model Number GIF-HQ190
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer returned the gastrointestinal videoscope to the service center for a separate issue.During the initial evaluation of the device, there was an e217 error (scope error).There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.The device evaluation found there was an e217 error (scope error).Based on the results of the investigation, the probable root cause was component failure, the printed circuit board was defective.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50933142
Report Number9610595-2026-80075
Device Sequence Number9597930
Product Code FDS
UDI-Device Identifier04953170436734
UDI-Public04953170436734
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K131780
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGIF-HQ190
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/09/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-