| Model Number |
3501 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
09/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) system delivered an inappropriate shock due to oversensing of noise.Technical services (ts) reviewed device data and suspected the noise was caused by myopotentials or electromagnetic interference (emi).Ts discussed troubleshooting options, including isometrics testing and programming to another vector.This s-icd system was in secondary vector during the shock episode.This electrode remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|