• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; METER, GLUCOSE, EVENCARE G2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; METER, GLUCOSE, EVENCARE G2 Back to Search Results
Catalog Number MPH1540
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "insert it but can not get a reading, change batteries but just goes off, first to months used received higher rating.".
 
Additional Manufacturer Narrative
The customer reported receiving a medline evencare g2 glucose meter that was shutting off during use, and the complaint text stated, "insert it but can not get a reading, change batteries but just goes off, first to months used received higher rating." at this time, a sample was not received for investigation; photos of the product were provided by the customer, but no determinations are able to be made based on the photos provided, and the investigation included review of recent complaint trends/scar activity for the past 12 months and vendor internal investigation, including dhr review and inspection records review, in which no abnormalities were found.The investigation results were unconfirmed, and the complaint could not be confirmed.Cause was "sample required to determine cause"; based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER, GLUCOSE, EVENCARE G2
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
pranay abbu
three lakes drive
northfield, IL 60093--275
MDR Report Key50933168
Report Number1417592-2026-53181
Device Sequence Number9734421
Product Code NBW
UDI-Device Identifier00884389102551
UDI-Public00884389102551
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K102037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH1540
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2024
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-