The customer reported receiving a medline evencare g2 glucose meter that was shutting off during use, and the complaint text stated, "insert it but can not get a reading, change batteries but just goes off, first to months used received higher rating." at this time, a sample was not received for investigation; photos of the product were provided by the customer, but no determinations are able to be made based on the photos provided, and the investigation included review of recent complaint trends/scar activity for the past 12 months and vendor internal investigation, including dhr review and inspection records review, in which no abnormalities were found.The investigation results were unconfirmed, and the complaint could not be confirmed.Cause was "sample required to determine cause"; based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|