Model Number 72400024
Medical Device Problem Codes Fluid/Blood Leak (1250); Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with an artificial urinary sphincter (aus) experienced recurrent urinary incontinence.The patient stated that the device stopped functioning.A surgical procedure was performed, during which a small hole was found at the neck of the balloon, which caused a fluid loss.As a result, the aus was removed.No further patient complications were reported.
 
Additional Manufacturer Narrative
Concomitant products implanted at the time of the event: model number/catalog number: 72404127, serial number: n/a, batch/lot number: 1100828095, model/catalog description: control pump fgs iz, unique identifier (udi)#: (b)(4), model number/catalog number: 720157-01, serial number: n/a, batch/lot number: 1100779757, model/catalog description: cuff fgs 3.5 cm inhibizone, unique identifier (udi)#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50933268
Report Number2124215-2026-51672
Device Sequence Number21223387
Product Code EZY
UDI-Device Identifier00878953000626
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72400024
Device Catalogue Number72400024
Device Lot Number1100668681
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured01/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
Patient SexUnknown
-
-