• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba CapSure Epicardial Lead; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba CapSure Epicardial Lead; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 4968-25
Medical Device Problem Code Under-Sensing (1661)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the right atrial (ra) lead and the right ventricular (rv) lead exhibited undersensing.The leads remain in use. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: w1dr01 ipg, implant date: (b)(6) 2025.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CapSure Epicardial Lead
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50933336
Report Number2649622-2026-28748
Device Sequence Number9734430
Product Code DTB
UDI-Device Identifier00613994492272
UDI-Public(01)00613994492272(17)180623
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P950024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/23/2018
Device Model Number4968-25
Device Catalogue Number4968-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/29/2026
Date Device Manufactured08/10/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4968-25 LEAD
Patient Age15 YR
Patient SexMale
-
-