It was reported that during a left atrial appendage (laa) procedure under general anesthesia, a versacross access solution was used for transseptal access.Baseline imaging with tee identified a pre-existing pericardial effusion measuring 0.7 cm, with fluid noted behind the laa and around the rv.A non-boston scientific mapping catheter was used intra-procedurally.The physician performed a transseptal puncture using an 8.5f fixed versacross sheath and versacross pigtail wire.After the wire was advanced into the laa and the sheath withdrawn, a non-boston scientific sheath was loaded over the wire.At that time, the ice representative noted a growing pericardial effusion, which was subsequently confirmed by tee.The patient became hemodynamically unstable with a drop in blood pressure, prompting emergent pericardiocentesis.Approximately 1,500 ml of bright red blood was evacuated and returned to the patient via cell saver autotransfusion.Due to continued blood re-accumulation, the patient was transferred to surgery, where an laa perforation attributed to the versacross wire was reportedly confirmed and managed with a pericardial window and laa clipping.On the following day, the patient was reported to be doing well, with planned extubation later that evening.The patient was not receiving antithrombotic therapy.The patient recovered very quickly and was discharged the same week of the event.The device was disposed.
|