| Model Number |
AP-03000H |
| Medical Device Problem Codes |
Material Integrity Problem (2978); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that there was no flow to the pump, resulting in additional surgery.The patient was implanted with an intera 3000 hepatic artery infusion pump.During a pump refill procedure on an unknown date, all 30 ml of medication was returned, indicating no flow.Imaging revealed that the catheter was twisted around a suture within the pump pocket.The surgeon opened the pump pocket and unraveled the catheter.The catheter was flushed using a special bolus needle to confirm that it was not occluded, and the pump was re-sutured into the pocket.The troubleshooting resolved the observed catheter twisting.The patient was scheduled for a hepatic arterial perfusion scintigraphy (haps) scan in a few days and a pump refill the following week.
|
| |
|
Additional Manufacturer Narrative
|
|
Section b3: date of event estimated using the date when boston scientific first became aware of the complaint.Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other specific product information.If additional details become available, a supplemental report will be submitted.
|
| |
|