| Model Number |
EZ-28V |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during insertion of an intraocular lens (iol) into the left eye (os), the surgeon heard a cracking sound from the injector, and the lens exited through the side of the injector rather than the tip.The injector containing the iol was withdrawn from the eye, and a backup lens was implanted.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was discarded by the reporting facility and was not available for evaluation.A photograph provided for review showed the intraocular lens (iol) exiting through the side of the inserter near the tip.A review of the device history record (dhr) did not identify any anomalies or nonconformities that could be related to this event.The lot history, trend analysis, risk analysis, and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause could not be conclusively determined.
|
| |
|