| Model Number |
7841 |
| Medical Device Problem Codes |
Failure to Capture (1081); High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range impedance measurements and a signal artifact monitoring (sam) episode due to artifact related to the minute ventilation (mv) feature signal on its right atrial (ra) lead.Additionally, there was loss of capture (loc) for its ra lead following the sam episode.Technical services (ts) was consulted and discussed stored data.The device will have its programming adjusted and the patient will continue to be monitored.This device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A good faith effort was performed to obtain additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
| |
|
Search Alerts/Recalls
|