| Model Number |
6372 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
06/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced a shocking sensation and was unable to enter their device into mri mode.Diagnostics revealed high impedances on bilateral extensions.As a result, surgical intervention was undertaken wherein the bilateral extensions were explanted and replaced to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|