| Model Number |
PT-002516 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the patient that the needle did not move forward when the pod was activated; this indicates a needle mechanism failure.The pod was worn for greater than 48 hours.No impact to the patient's glucose level was reported.As treatment, a new pod was placed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed, and the product lot met all acceptance criteria.Locked down smartphone: lockdown
omnipod software app version: 4.0.2
operating system: n5004l-am-q-mv01602-06-01.06
hardware: n5004l
cgm sensor type: data not available
please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|