The customer reported that in 2 separate cases the scalpel blade fractured during use for incision, and stated no adverse event occurred after the scalpel broke and no medical or surgical treatment was required.One box of brand-new items in their original packaging was received for investigation; visual and physical inspections were conducted, no visual abnormalities were observed, and during physical inspection the blade fractured and broke in half while handling and gently bending the item for closer evaluation.Based on the complaint details, sample evaluation, and investigation, the complaint has been confirmed.Physical inspection revealed that the item exhibited reduced mechanical integrity and broke easily; the root cause of this issue cannot be determined.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|