| Model Number |
PT-002516 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient's blood glucose level rose to greater than 250 mg/dl, while wearing the pod between 4 and 24 hours.The pod reportedly detached from the arm, causing the cannula to dislodge.Details regarding treatment were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the complainant the device will not be returned for investigation.It was reported that the cannula dislodged from the infusion site.We are unable to confirm or determine the root cause of the reported dislodged cannula and if it could have contributed to the reported hyperglycemia.Lot release was found to have met all acceptance criteria.Locked down smartphone: lockdown
omnipod software app version: 4.0.2
operating system: n5004l-am-q-mv01602-06-01.06
hardware: n5004l
cgm sensor type: l2+
please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|