Tandem Diabetes Care Tandem Mobi Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number 1004000
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that occlusion alarms occurred on mobi pump over the past week, and technical support was unable to verify alarm number.There was no adverse impact to the customer¿s blood glucose level.Customer changed infusion set and cartridge and resumed insulin, and due to ongoing occlusion issues despite changing all supplies, technical support arranged replacement pump, confirmed shipping details, instructed customer to use injected insulin as backup therapy, provided instructions to place pump in storage mode and return problematic pump within 10 days, and advised customer to enter pump settings and reconnect cgm on replacement pump.
 
Additional Manufacturer Narrative
In use at the time of the event: cgm model: unknown.Mobile os: unknown / unknown / unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Tandem Mobi Insulin Pump with Control-IQ+ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50933894
Report Number3013756811-2026-176892
Device Sequence Number9689023
Product Code QFG
UDI-Device Identifier00389152370019
UDI-Public(01)00389152370019(22)1013701(11)240822
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K240309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1004000
Device Catalogue Number1013700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/22/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Lispro - Insulin; Tandem - Cartridge; Unomedical - Infusion Set
Patient Age3 YR
Patient SexFemale
-
-