• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL Manufacturing Corporation LifeVest WCD 4000 System; Wearable Cardioverter Defibrillator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL Manufacturing Corporation LifeVest WCD 4000 System; Wearable Cardioverter Defibrillator Back to Search Results
Model Number LifeVest WCD 4000 System
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
A us distributor reported that a patient was experiencing check therapy electrode messages.
 
Additional Manufacturer Narrative
Device evaluation of electrode belt has been completed.The reported problem (tes non-functional) has been confirmed.Upon investigation the electrode belt failed a therapy electrode recognition test.The cause for the failure was isolated to an open pulse wire in the cable connecting ecg "a" and the front therapy electrode.The root cause for the open wire was unable to be identified.There was no adverse event that resulted from the damaged belt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LifeVest WCD 4000 System
Common Device Name
Wearable Cardioverter Defibrillator
Manufacturer (Section D)
ZOLL Manufacturing Corporation
121 gamma drive
pittsburgh PA 15238
Manufacturer (Section G)
ZOLL Manufacturing Corporation
121 gamma drive
pittsburgh PA 15238
Manufacturer Contact
zoll mfg corporation
121 gamma drive
pittsburgh, PA 15238
MDR Report Key50933968
Report Number3008642652-2026-09462
Device Sequence Number21519996
Product Code MVK
UDI-Device Identifier00855778005135
UDI-Public(01)00855778005135(17)270918
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberLifeVest WCD 4000 System
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2026
Initial Report FDA Received Date09/29/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-