| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a patient was experiencing check therapy electrode messages.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of electrode belt has been completed.The reported problem (tes non-functional) has been confirmed.Upon investigation the electrode belt failed a therapy electrode recognition test.The cause for the failure was isolated to an open pulse wire in the cable connecting ecg "a" and the front therapy electrode.The root cause for the open wire was unable to be identified.There was no adverse event that resulted from the damaged belt.
|
| |
|