Clinical staff was having issues on the caps not staying secured.Medline received 50ea hypodermic safety needle in new condition and tested all 50ea samples; the caps stayed firmly on the needle and showed no signs of falling off during the opening process or when being attached to the syringe.The complaint was not confirmed, and the cause was not confirmed with received sample.Material trending via sap showed no similar complaints during the last 12 months; the samples worked as intended and medline was unable to confirm that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|