| Catalog Number |
8065000478 |
| Medical Device Problem Codes |
Suction Problem (2170); Failure to Cut (2587)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A healthcare professional reported that the vitrectome malfunctioned during surgery, as it failed to perform both cutting and vacuum functions.The procedure type was unknown.Even after reducing the cutting speed to 10000 cuts per minute (cpm) and changing the pump of the console the issue was not resolved.The procedure was able to continue after the affected vitrectome was replaced with a new one.It was unknown how the surgery was completed.There was no information about the patient impact.Additional information could not be obtained because the initial reporter declined further follow-up contact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|