| Model Number |
WA50052A |
| Medical Device Problem Codes |
Fracture (1260); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the rigid video laparoscope had kink in the optics during reprocessing.During the initial evaluation of the device it was observed, fiber bonding in the outer tube area (distal end) is damaged and charged coupled device (r-unit) has a kink noted or observed.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.No additional reportable investigation findings.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the outer tube is deformed.The most probable cause is traced to component failure.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|