| Model Number |
MX40 1.4 GHz Smart Hopping |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on a mx40 1.4 ghz smart hopping monitor indicating that there is a speaker malfunction on the mx40.The device was reported to be in clinical use.No adverse event to the patient or user was reported.The staff used another device to monitor the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by bench repair personnel.During device evaluation, evidence of tampering was identified.Specifically, the front bezel/display assembly and rear housing were determined to be non-genuine philips components.Due to the presence of non-genuine parts and evidence of tampering, the device was returned to the customer unrepaired.No additional investigation or functional testing was performed.At this time, it is unknown whether the device generated audible sound during the reported event.Additional information has been requested from the customer.Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
| |
|