| Model Number |
D143 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the health care professional (hcp) called technical services (ts) for a consult review of the implantable cardioverter defibrillator (icd) stored episode due to concerns of oversensing noise.The hcp stated that the episode occurred while the patient was exercising.Upon review, the presenting electrogram (egm) showed oversensed noise that appears to be inhibiting pacing.Its noted patient is not pacing dependent.Ts also noted a drop in right ventricular (rv) lead pacing impedance from 900 to 600ohms which are still considered to be within normal limits (wnl).Ts recommended an in clinic visit for isometric evaluation of the lead.Subsequently, a revision procedure was performed.The icd was explanted and replaced with a new device.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|