| Model Number |
PT-001444 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Difficult to Insert (1316); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the legal guardian (lg) that the patient's blood glucose level rose to 310-320 mg/dl while wearing the pod between 36 and 48 hours despite boluses.The lg reported smelling insulin and the adhesive was pulling off.The lg reported being unsure if the cannula was properly seated in the infusion site and when removed, the pod's cannula was found bent.As treatment, the lg planned to apply a new pod.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned; however, the investigation is not yet complete.When the investigation is complete, a supplemental report will be submitted with the investigation results.Lot release records were reviewed, and the product lot met all acceptance criteria.Locked down smartphone: lockdown.Omnipod software app version: 4.0.2.Operating system: n5004l-am-q-mv01602-06-01.06.Hardware: n5004l.Cgm sensor type: freestyle libre 2 plus.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|