| Model Number |
TRB24-REG-01 |
| Medical Device Problem Codes |
Gas/Air Leak (2946); Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Terumo medical corporation received the following information: it was reported that the tr band was noted to be losing air approximately forty-five minutes after completion of the procedure while the patient was in the post-operative area.When the tr band was removed off the patient, they inflated the balloons and when pressure was put on the balloons, they would deflate.There was no patient injury.Manual pressure was applied and hemostasis achieved.There was no blood loss.The patient remained stable.A total of 18 ml of air was injected into the tr band when it was initially applied to the patient¿s wrist.The tr band is typically left in place for sixty minutes before air titration begins for patients who receive less than fifty units of heparin, and for one hundred and twenty minutes for patients who receive more than fifty units of heparin.In this case, all of the air had already been lost from the tr band when titration was initiated in the post-operative area.There was no noticeable damage to the device, and the device had not been manipulated in any way prior to use or during storage.Before use, the tr band was stored in its original tr band box on enclosed shelves.The type of procedure performed prior to the use of the tr band was left heart catheterization (lhc).
|
| |
|
Additional Manufacturer Narrative
|
|
A1: patient identifier: requested, not provided due to hospital policy
a2: age & date of birth: requested, not provided due to hospital policy
a3a: sex: requested, not provided due to hospital policy
a4: weight: requested, not provided due to hospital policy
a5: ethnicity: requested, not provided due to hospital policy
a6: race: requested, not provided due to hospital policy
d6a: implanted date: device was not implanted
d6b: explanted date: device was not explanted
e3: occupation: (b)(6).The actual device has been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.
|
| |
|