• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Biosense Webster Inc VARIPULSE Bi-Directional Catheter; Percutaneous Cardiac Abl Cath for Treatmt of AFib w Irreversible Electroporation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Biosense Webster Inc VARIPULSE Bi-Directional Catheter; Percutaneous Cardiac Abl Cath for Treatmt of AFib w Irreversible Electroporation Back to Search Results
Catalog Number D141201
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 08/31/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a varipulse bi-directional catheter and the patient experienced a cardiac tamponade which required pericardiocentesis.Transseptal puncture was performed with a radiofrequency (rf) needle (boston scientifics).A total of 16 ablations by pulse field ablation was done.The procedure was completed without issue.After sheath removal, the patient's blood pressure decreased, and a cardiac tamponade was discovered.Pericardial drainage was performed, and the blood pressure subsequently stabilized.The patient was under observation and has fully recovered.Physician's judgment on health hazard is that it was non-serious (moderate/minor).The physician's comment regarding relationship between the event and the product is that as the varipulse catheter was used only within the left atrium and the drained blood was venous blood; the issue was considered unlikely to be related to the device.The physician believes it was procedure related.There were no abnormalities observed prior to or during product use.
 
Additional Manufacturer Narrative
Device investigation details: an analysis of the product could not be performed since a physical sample was not received for evaluation.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device number lot 31888693l and no internal actions related to the complaint were found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VARIPULSE Bi-Directional Catheter
Common Device Name
Percutaneous Cardiac Abl Cath for Treatmt of AFib w Irreversible Electroporation
Manufacturer (Section D)
Biosense Webster Inc
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
Biosense Webster Inc (Juarez)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
katie seppa
31 technology dr
irvine, CA 92618
+1(513)337-8052
MDR Report Key50935392
Report Number3013300026-2026-03196
Device Sequence Number10734839
Product Code QZI
UDI-Device Identifier10846835025460
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P240006
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberD141201
Device Lot Number31888693L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/04/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening;
Patient SexUnknown
Patient RaceAsian
-
-