This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
investigation summary: the product was not returned to depuy synthes for evaluation, however a photo was provided for evaluation.Visual inspection of the returned photo found that the needle is shown in used condition.The needle was found broken in half, traces of being deformed and forced.Ductile fracture of metal can be observed.The overall complaint was confirmed as the observed condition of the omnispan meniscal repair 12deg would have contributed to the complained device issue.¿
based on the investigation findings, the potential cause is traced to a mechanical overload that was applied on the needle leading to a ductile fracture; as per the instructions for use (ifu); the proper steps for manipulation of the device are provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|