| Model Number |
DDU-100A |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that their aed's asi is flashing red, and the device is reporting a service code.The customer did not report this occurring during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the aed's log files identified that the customer was performing excessive self-testing of the aed which reduced the previous battery's life expectancy.Troubleshooting of the aed with the customer identified that once the new battery was install and a manual self test run the aed functioned as designed and could stay in service.As a follow up with the customer, the proper maintenance of this device was reviewed.
|
| |
|
Search Alerts/Recalls
|