MAUDE Adverse Event Report: Cook Inc ULTRATHANE DAWSON-MUELLER MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; FGE; Stents, Drains and Dilators for the Biliary Ducts

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Cook Inc ULTRATHANE DAWSON-MUELLER MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; FGE; Stents, Drains and Dilators for the Biliary Ducts Back to Search Results
Model Number G13768
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stiffening cannula could not be removed from an ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter during an unspecified procedure.The drainage catheter with stiffener were advanced over the wire guide; however, when attempting to remove the stiffening cannula, it was found to be firmly lodged in the catheter and could not be removed.Another device was subsequently used, but the stiffening cannula of that device also could not be removed.The procedure was completed using an alternative like device from a different catalog number.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.D2a- additional common names: gbo catheter, nephrostomy, general & plastic surgery; lje catheter, nephrostomy d2b- additional product codes: gbo; lje h3 - the device is not being returned for evaluation.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE DAWSON-MUELLER MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Common Device Name
FGE; Stents, Drains and Dilators for the Biliary Ducts
Manufacturer (Section D)
Cook Inc
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
+1(812)339-2235
MDR Report Key50935629
Report Number1820334-2026-01010
Device Sequence Number9509254
Product Code FGE
UDI-Device Identifier00827002137684
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG13768
Device Catalogue NumberULT7.0-35-25-P-5S-CLDM-WF-HC
Device Lot Number17157682
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2026
Initial Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/04/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-