| Model Number |
G13768 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the stiffening cannula could not be removed from an ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter during an unspecified procedure.The drainage catheter with stiffener were advanced over the wire guide; however, when attempting to remove the stiffening cannula, it was found to be firmly lodged in the catheter and could not be removed.Another device was subsequently used, but the stiffening cannula of that device also could not be removed.The procedure was completed using an alternative like device from a different catalog number.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown or unavailable.D2a- additional common names: gbo catheter, nephrostomy, general & plastic surgery; lje catheter, nephrostomy
d2b- additional product codes: gbo; lje
h3 - the device is not being returned for evaluation.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|