The customer reported the head motor gears are stripped.Medline received one bed motor only in used condition, and when bench tested with known working components, the motor was running, but the drive shaft does not turn; there was no physical damage to the motor case, power cord, or plug end.The quality complaint was confirmed, and the issue may be due to internal stripped gears; the cause was documented as cannot determine cause with received sample.The division will monitor through quarterly trending, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|